Model 2461 Interactive SourceMeter® Instrument Reference Manual

smu.contact.threshold

This attribute sets the threshold value for contact resistance for the contact check status functions.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
smu.contact.threshold = level
level = smu.contact.threshold
level
Details
The contact check functions indicate a failure when the connection path resistance exceeds the value
specified by this command.
This command can be sent when contact check isn't enabled. However, you must enable contact
check before checking the status of connections.
Example
smu.contact.threshold = smu.THRESHOLD_15_OHM
print(smu.contact.threshold)
Also see
smu.contact.check()
smu.contact.checkall()
smu.contact.enable
2461-901-01 A/November 2015
Affected by
Instrument reset
Power cycle
The level of threshold resistance above which the contact check functions indicate a
failure:
2 Ω: smu.THRESHOLD_2_OHM
15 Ω: smu.THRESHOLD_15_OHM
50 Ω: smu.THRESHOLD_50_OHM
(on page 8-102)
(on page 8-103)
(on page 8-104)
Section 8: TSP command reference
Where saved
Not saved
Set the contact check function to indicate a
failure if resistance is greater than 15 Ω.
Query the contact check threshold setting.
Output:
smu.THRESHOLD_15_OHM
Indicates that the contact check threshold is
set to 15 Ω.
Default value
smu.THRESHOLD_50_OHM
8-105