Output[1]:Interlock:tripped; Output[1][:State] - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual

:OUTPut[1]:INTerlock:TRIPped?

This command indicates that the interlock has been tripped.
Type
Query only
Usage
:OUTPut[1]:INTerlock:TRIPped?
Details
This command gives you the status of the interlock. When the safety interlock signal is asserted, all
voltage ranges of the instrument are available. However, when the safety interlock signal is not
asserted, the V range is disabled, limiting the nominal output to less than ±42 V.
When the interlock is not asserted:
If 1 is returned, the interlock signal is asserted and all voltage ranges are available.
If 0 is returned, the interlock is not asserted and the 100 V range is disabled. Lower voltage ranges
are available.
Example
OUTP:INT:TRIP?
Also see
None

:OUTPut[1][:STATe]

This command enables or disables the source output.
Type
Command and query
Usage
:OUTPut[1][:STATe] <state>
:OUTPut[1][:STATe]?
<state>
2461-901-01 A/November 2015
Affected by
Not applicable
The front-panel INTERLOCK indicator is off.
High voltage ranges are disabled.
If you attempt to turn on the source with a voltage more than ±21 V, an event message is generated.
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
Where saved
Not applicable
If the interlock is not asserted, returns 0.
If the interlock is asserted, returns 1.
Where saved
Save settings
Source configuration list
Turn source off: 0 or OFF
Turn source on: 1 or ON
Section 6: SCPI command reference
Default value
Not applicable
Default value
0 (OFF)
6-49