Acal.lastrun.tempdiff - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 8: TSP command reference
Example
print(acal.lastrun.internaltemp)
Also see

acal.lastrun.tempdiff

acal.run()
(on page 8-9)
acal.lastrun.tempdiff
This attribute returns the difference between the internal temperature and the temperature when autocalibration
was last run.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
temperature = acal.lastrun.tempdiff
temperature
Details
The temperature is displayed in Celsius (°C).
Example
print(acal.lastrun.tempdiff)
Also see
acal.lastrun.internaltemp
acal.run()
(on page 8-9)
8-8
Model 2461 Interactive SourceMeter® Instrument Reference Manual
(on page 8-8)
Affected by
Not applicable
The internal temperature
(on page 8-7)
Returns the internal temperature of the instrument when
auto calibration was last run.
Example output:
63.167084
Where saved
Not applicable
Returns the difference between the temperature of the
instrument when auto calibration was run and the present
internal temperature.
Example output:
4.5678
Default value
Not applicable
2461-901-01 A/November 2015