Section 8: TSP command reference
Also see
Configuration lists
Pulse operation
smu.source.sweeplist()
trigger.model.initiate()
Trigger model
smu.source.pulsesweeplog()
This function sets up a logarithmic pulse sweep for a set number of source points.
Type
TSP-Link accessible
Function
Yes
Usage
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual)
smu.source.pulsesweeplog(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual, asymptote)
configListName
biasLevel
start
8-202
(on page 3-30)
(on page 3-64)
(on page 8-214)
(on page 8-259)
(on page 3-107)
Affected by
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
Current: -7.35 A to 7.35 A
•
Voltage: -105 V to 105 V
•
The voltage or current source level at which the pulse sweep starts:
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
•
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Where saved
Default value
2461-901-01 A/November 2015