Status Model; Standard Event Register - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register.
Bit
0
1
2
3
4
5
6
7
Command errors include:
Constants are only available if you are using the TSP command set. If you are using the SCPI
command set, you must use the decimal values.
Example
*ESE 129
Also see
*CLS
(on page B-2)

Standard Event Register

Status model

2461-901-01 A/November 2015
Decimal
Constant
value
1
status.standard.OPC
2
Not used
4
status.standard.QYE
8
Not used
16
Not used
32
Not used
64
Not used
128
status.standard.PON
IEEE Std 488.2 syntax error: The instrument received a message that does not follow the defined
syntax of the IEEE Std 488.2 standard.
Semantic error: The instrument received a command that was misspelled or received an optional IEEE
Std 488.2 command that is not implemented in the instrument.
GET error: The instrument received a Group Execute Trigger (GET) inside a program message.
(on page C-3)
(on page C-1)
When set, indicates the following has occurred:
All pending selected instrument operations are
complete and the instrument is ready to accept new
commands. The bit is set in response to an
(on page B-6) command or TSP
92) function.
Not used.
Attempt to read data from an empty Output Queue.
Not used.
Not used.
Not used.
Not used.
The instrument has been turned off and turned back
on since the last time this register was read.
*ESE 129 sets the Status Enable register of the Standard Event
Register to binary 10000001, which enables the PON and OPC bits.
Appendix B: Common commands
*OPC
opc()
(on page 8-
B-3