Source[1]:<Function>:Protection[:Level]:Tripped; Source[1]:<Function>:Range - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
:SOURce[1]:<function>:PROTection[:LEVel]:TRIPped?
This command indicates if the overvoltage source protection feature is active.
Type
Query only
Usage
:SOURce[1]:VOLTage:PROTection[:LEVel]:TRIPped?
Details
When overvoltage protection is active, the instrument restricts the maximum voltage level that the
instrument can source.
If the voltage source does not exceed the set limits, the return is 0. If the voltage source exceeds the
set limits, the return is 1.
Example
SOUR:VOLT:PROT:TRIP?
Also see
Overvoltage protection
:SOURce[1]:<function>:PROTection[:LEVel]
:SOURce[1]:<function>:RANGe
This command selects the range for the source for the selected source function.
Type
Command and query
Usage
:SOURce[1]:<function>:RANGe <n>
:SOURce[1]:<function>:RANGe?
:SOURce[1]:<function>:RANGe? DEFault
:SOURce[1]:<function>:RANGe? MINimum
:SOURce[1]:<function>:RANGe? MAXimum
<function>
<n>
2461-901-01 A/November 2015
Affected by
Not applicable
(on page 2-121)
(on page 6-96)
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
The source function to which this setting applies:
Current: CURRent
Voltage: VOLTage
Set to the maximum expected voltage or current to be sourced; see
Details for values; the ranges are:
Current: -10 A to 10 A
Voltage: -100 V to 100 V
Section 6: SCPI command reference
Where saved
Not applicable
If overvoltage protection is active, the output is:
1
Where saved
Save settings
Source configuration list
Default value
Not applicable
Default value
Current: 1 µA
Voltage: 200 mV
6-97