smu.source.sweeplog()

This function sets up a logarithmic sweep for a set number of measurement points.
Type
TSP-Link accessible
Function
Yes
Usage
smu.source.sweeplog(configListName, start, stop, points)
smu.source.sweeplog(configListName, start, stop, points, sDelay)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count, rangeType)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count, rangeType,
failAbort)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count, rangeType,
failAbort, dual)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count, rangeType,
failAbort, dual, bufferName)
smu.source.sweeplog(configListName, start, stop, points, sDelay, count, rangeType,
failAbort, dual, bufferName, asymptote)
configListName
start
stop
points
sDelay
count
rangeType
failAbort
dual
bufferName
asymptote
8-216
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Where saved
A string that contains the name of the configuration list that the instrument will
create for this sweep
The voltage or current source level at which the sweep starts:
Current: 1 µA to 7.35 A
Voltage 200 mV to 105 V
The voltage or current at which the sweep stops:
Current: 1 µA to 7.35 A
Voltage 200 mV to 105 V
The number of source-measure points between the start and stop values of the
sweep (2 to 1e6); to calculate the number of source-measure points in a sweep, use
the following formula:
Points = [(Stop - Start) / Step] + 1
The delay between measurement points; default is smu.DELAY_AUTO, which
enables autodelay, or a specific delay value from 50 µs to 10 ks, or 0 for no delay
The number of times to run the sweep; default is 1:
Infinite loop: smu.INFINITE
Finite loop: 1 to 268,435,455
The source range that is used for the sweep:
Most sensitive source range for each source level in the sweep:
smu.RANGE_AUTO
Best fixed range: smu.RANGE_BEST (default)
Present source range for the entire sweep: smu.RANGE_FIXED
Complete the sweep if the source limit is exceeded: smu.OFF
Abort the sweep if the source limit is exceeded: smu.ON (default)
Determines if the sweep runs from start to stop and then from stop to start:
Sweep from start to stop only: smu.OFF (default)
Sweep from start to stop, then stop to start: smu.ON
Indicates the reading buffer to use; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Default is 0; see Details
Default value
2461-901-01 A/November 2015