[:Sense[1]]:<Function>:Azero[:State] - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
Also see
Filtering measurement data
[:SENSe[1]]:<function>:AVERage:COUNt
[:SENSe[1]]:<function>:AVERage[:STATe]
[:SENSe[1]]:<function>:AZERo[:STATe]
This command enables or disables automatic updates to the internal reference measurements (autozero) of the
instrument.
Type
Command and query
Usage
[:SENSe[1]]:<function>:AZERo[:STATe] <state>
[:SENSe[1]]:<function>:AZERo[:STATe]?
<function>
<state>
2461-901-01 A/November 2015
(on page 4-23)
(on page 6-53)
(on page 6-55)
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measure function:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
The status of autozero:
Disable autozero: OFF or 0
Enable autozero: ON or 1
Section 6: SCPI command reference
Set the averaging filter type to moving average,
with a filter count of 10.
Enable the averaging filter.
Set the averaging filter type to moving average,
with a filter count of 10.
Enable the averaging filter.
For voltage measurements, set the averaging
filter type to moving average, with a filter count
of 10.
Enable the averaging filter.
Where saved
Save settings
Measure configuration list
Default value
ON (1)
6-57