Trigger:block:source:state - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference

:TRIGger:BLOCk:SOURce:STATe

This command defines a trigger block that turns the output source on or off.
Type
Command only
Usage
:TRIGger:BLOCk:SOURce:STATe <blockNumber>, <state>
<blockNumber>
<state>
Details
The source output block determines if the output source is turned on or off when the trigger model
reaches this block.
This block does not determine the settings of the output source (such as the output voltage level and
source delay). The source settings are determined by either the present settings of the instrument or
by a source configuration list.
When you list trigger blocks, this block is listed as SOURCE_OUTPUT.
Example
TRIG:BLOC:SOUR:STAT 1, 1
TRIG:BLOC:DEL:CONS 2, 0.01
TRIG:BLOC:MEAS 3
TRIG:BLOC:BRAN:COUN 4, 100, 2
TRIG:BLOC:SOUR:STAT 5, 0
Also see
Wait block
(on page 3-110)
6-208
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
Disable the source: OFF
Enable the source: ON
Where saved
Save settings
This example turns the output on.
Delay 10 ms.
Make a measurement.
Loop and take 100 readings.
The output is turned off after the 100 readings
are made.
Default value
Not applicable
2461-901-01 A/November 2015