Keithley 2461 Reference Manual page 135

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 2: General operation
Model 2461 Interactive SourceMeter® Instrument Reference Manual
The front-panel display will look similar to the following example.
Figure 68: Resistance measurement SIMV SCPI example
Using TSP commands:
Send the following code:
reset()
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.autorange = smu.ON
smu.measure.unit = smu.UNIT_OHM
smu.measure.count = 5
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.level = 5e-6
smu.source.vlimit.level = 10
smu.source.output = smu.ON
smu.measure.read(defbuffer1)
for i = 1, defbuffer1.n do
print(defbuffer1.relativetimestamps[i], defbuffer1[i])
end
smu.source.output = smu.OFF
The front-panel display will look similar to the following example.
Figure 69: Resistance measurement SIMV and read ohms
2-108
2461-901-01 A/November 2015

Advertisement

Table of Contents
loading

Table of Contents