Trigger.digout[N].Release() - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 8: TSP command reference
Details
Setting the pulse width to zero (0) seconds asserts the trigger indefinitely. To release the trigger line,
use trigger.digout[N].release().
Example
digio.line[4].mode = digio.MODE_TRIGGER_OUT
trigger.digout[4].pulsewidth = 20e-6
Also see
digio.line[N].mode
digio.line[N].reset()
Digital I/O port configuration
trigger.digout[N].assert()

trigger.digout[N].release()

trigger.digout[N].release()
This function releases an indefinite length or latched trigger.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.digout[N].release()
N
Details
Releases a trigger that was asserted with an indefinite pulsewidth time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.line[4].mode = digio.MODE_TRIGGER_OUT
trigger.digout[4].release()
Also see
digio.line[N].mode
Digital I/O port configuration
trigger.digout[N].assert()
trigger.digout[N].pulsewidth
8-246
Model 2461 Interactive SourceMeter® Instrument Reference Manual
(on page 8-50)
(on page 8-52)
(on page 3-87)
(on page 8-244)
(on page 8-246)
Affected by
Digital I/O trigger line (1 to 6)
(on page 8-50)
(on page 3-87)
(on page 8-244)
(on page 8-245)
Sets the pulse width for trigger line 4 to
20 µs.
Where saved
Releases digital I/O trigger line 4.