Source[1]:Sweep:<Function>:Log - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference
:SOURce[1]:SWEep:<function>:LOG
This command sets up a logarithmic sweep for a set number of measurement points.
Type
Command only
Usage
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>,
<rangeType>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>,
<rangeType>, <failAbort>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>,
<rangeType>, <failAbort>, <dual>
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>,
<rangeType>, <failAbort>, <dual>, "<bufferName>"
:SOURce[1]:SWEep:<function>:LOG <start>, <stop>, <points>, <delay>, <count>,
<rangeType>, <failAbort>, <dual>, "<bufferName>", <asymptote>
<function>
<start>
<stop>
<points>
<delay>
<count>
<rangeType>
<failAbort>
<dual>
6-130
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The source function:
CURRent
VOLTage
The voltage or current source level at which the sweep starts:
Current: 1 µA to 7.35 A
Voltage: 200 mV to 105 V
The voltage or current at which the sweep stops:
Current: 1 µA to 7.35 A
Voltage: 200 mV to 105 V
The number of source-measure points between the start and stop values of the
sweep (2 to 1e6); to calculate the number of source-measure points in a sweep, use
the following formula:
Points = [(Stop - Start) / Step] + 1
The delay between measurement points; default is -1, which enables autodelay, or
a specific delay value from 50 µs to 10,000 s, or 0 for no delay
The number of times to run the sweep; default is 1:
Infinite loop: 0
Finite loop: 1 to 268435455
The source range that is used for the sweep:
Most sensitive source range for each source level in the sweep: AUTO
Best fixed range: BEST (default)
Present source range for the entire sweep: FIXed
Abort the sweep if the source limit is exceeded: ON (default)
Complete the sweep if the source limit is exceeded: OFF
Determines if the sweep runs from start to stop and then from stop to start:
Sweep from start to stop only: OFF (default)
Sweep from start to stop, then stop to start: ON
Where saved
Save settings
Default value
Not applicable