Trace:actual:end - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
Details
This command returns the number of readings stored in the buffer.
Example
TRACe:MAKE "testData", 200
COUN 10
MEASure:CURRent? "testData"
:TRACe:ACTual?
:TRACe:ACTual? "testData"
Also see
Reading buffers
Remote buffer operation
:TRACe:MAKE

:TRACe:ACTual:END?

This command indicates the last index in a reading buffer.
Type
Query only
Usage
:TRACe:ACTual:END?
:TRACe:ACTual:END? "<bufferName>"
<bufferName>
Details
Use this command to find the ending index in a reading buffer.
2461-901-01 A/November 2015
(on page 3-2)
(on page 3-23)
(on page 6-165)
Affected by
Recall settings
Instrument reset
Power cycle
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Creates 200 element reading buffer named testData.
Set the measurement count to 10.
Set the measurement function to current. Make readings, and
store the readings in testData. Returns the 10
measurement reading after taking all 10 readings.
Returns the number of readings in defbuffer1.
Example output:
850
Returns the number of readings in the buffer testData.
Example output:
10
Where saved
Not applicable
Section 6: SCPI command reference
th
Default value
Not applicable
6-157