:TRIGger:DIGital<n>:IN:CLEar
This command clears the trigger event on a digital input line.
Type
Command only
Usage
:TRIGger:DIGital<n>:IN:CLEar
<n>
Details
The event detector of a trigger enters the detected state when an event is detected. For the specified
trigger line, this command clears the event detector, discards the history, and clears the overrun
status(sets the overrun status to 0).
For this block to work as expected, make sure you configure the trigger type and line state of the
digital line for use with the trigger model (use the digital line mode command).
Example
:TRIG:DIG2:IN:CLE
Also see
:DIGital:LINE<n>:MODE
Digital I/O port configuration
:TRIGger:DIGital<n>:IN:OVERrun?
:TRIGger:DIGital<n>:IN:EDGE
This command sets the edge used by the trigger event detector on the given trigger line.
Type
Command and query
Usage
:TRIGger:DIGital<n>:IN:EDGE <detectedEdge>
:TRIGger:DIGital<n>:IN:EDGE?
<n>
<detectedEdge>
2461-901-01 A/November 2015
Affected by
Not applicable
Digital I/O trigger line (1 to 6)
(on page 6-35)
(on page 3-87)
(on page 6-212)
Affected by
Recall settings
Instrument reset
Power cycle
Digital I/O trigger line (1 to 6)
The trigger edge value:
Detect falling-edge triggers as inputs: FALLing
Detect rising-edge triggers as inputs: RISing
Detect either falling or rising-edge triggers as inputs: EITHer
See Details for descriptions of values
Section 6: SCPI command reference
Where saved
Not applicable
Clears the trigger event detector on I/O line 2.
Where saved
Save settings
Default value
Not applicable
Default value
FALL