Trigger:block:delay:constant - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual

:TRIGger:BLOCk:DELay:CONStant

This command adds a constant delay to the trigger model.
Type
Command only
Usage
:TRIGger:BLOCk:DELay:CONStant <blockNumber>, <time>
<blockNumber>
<time>
Details
When trigger model execution reaches a delay block, it stops normal measurement and trigger model
operation for the amount of time set by the delay. Background measurements continue to be made,
and if any previously executed block started infinite measurements, they also continue to be made.
This delay waits for the set amount of delay time to elapse before proceeding to the next block in the
trigger model.
If other delays have been set, this delay is in addition to the other delays.
Example
SOUR:CONF:LIST:CRE "ampLevel"
SOUR:CONF:LIST:CRE "biasLevel"
SOUR:FUNC VOLT
SENS:FUNC "CURR"
SOUR:VOLT:LEV 5
SOUR:CONF:LIST:STORE "ampLevel"
SOUR:VOLT:LEV 0
SOUR:CONF:LIST:STORE "biasLevel"
TRIG:BLOC:SOUR:STATE 1, ON
TRIG:BLOCK:CONF:RECALL 2, "ampLevel", 1
TRIG:BLOC:DEL:CONS 3, 0.1
TRIG:BLOCK:MEAS 4
TRIG:BLOCK:CONF:RECALL 5, "biasLevel", 1
TRIG:BLOCK:DEL:CONS 6, 0.2
TRIG:BLOCK:BRAN:COUN 7,19,2
INIT
Also see
None
2461-901-01 A/November 2015
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The amount of time to delay (167 ns to 10,000 s, or 0 for no delay)
Section 6: SCPI command reference
Where saved
Save settings
Create configuration lists named ampLevel and
biasLevel.
Set the source function to 5 V and the
measurement function to current.
Store the settings in the ampLevel configuration
list at index 1.
Set the voltage level to 0 V.
Store the setting in the biasLevel configuration
list at index 1.
Set block 1 to turn the source output on.
Set block 2 to recall the ampLevel settings from
index 1.
Set block 3 to provide a constant delay of 0.1 s.
Set block 4 to make a measurement.
Set block 5 to recall the biasLevel settings at
index 1.
Set block 6 to provide a constant delay of 0.2 s.
Set block 7 to repeat block 2 nineteen times.
Default value
Not applicable
6-199

Advertisement

Table of Contents
loading

Table of Contents