Model 2461 Interactive SourceMeter® Instrument Reference Manual
[:SENSe[1]]:<function>:RELative
This command contains the relative offset value.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RELative <n>
[:SENSe[1]]:<function>:RELative?
[:SENSe[1]]:<function>:RELative? DEFault
[:SENSe[1]]:<function>:RELative? MINimum
[:SENSe[1]]:<function>:RELative? MAXimum
<function>
<n>
Details
This command specifies the relative offset value that can be applied to new measurements. When
relative offset is enabled, all subsequent measured readings are offset by the value that is set for this
command.
You can set this value, or have the instrument acquire a value. If the instrument acquires the value,
read this setting to return the value that was measured internally.
Example
CURR:REL .5
CURR:REL:STAT ON
Also see
Relative offset
[:SENSe[1]]:<function>:RELative:ACQuire
[:SENSe[1]]:<function>:RELative:STATe
2461-901-01 A/November 2015
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measure function:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
The digitize function:
Digitize current: DIGitize:CURRent
Digitize voltage: DIGitize:VOLTage
The relative offset value:
Current: −10 to 10
Resistance: –2e8 to 2e8
Voltage: −100 to 100
Digitize current: −10 to 10
Digitize voltage: −100 to 100
(on page 3-79)
(on page 6-69)
Where saved
Save settings
Measure configuration list
Set the relative offset for current measurements to 0.5.
Enable relative offset.
(on page 6-68)
Section 6: SCPI command reference
Default value
0
6-67