Trigger:block:config:recall - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference

:TRIGger:BLOCk:CONFig:RECall

This command recalls the system settings that are stored in a source or measure configuration list.
Type
Command only
Usage
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>"
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>", <index>
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>", <index>,
"<configurationList2>", <index2>
<blockNumber>
<configurationList>
<index>
<configurationList2>
<index2>
Details
When the trigger model reaches a configuration recall block, the settings in the specified configuration
list are recalled if a single configuration list is specified. If both measure and source configuration lists
are specified, measure and source settings are recalled from the next index in each list when this
block is reached. The index numbers recalled may not match; it depends on the number of indexes in
each list and what index number each list is on.
You can restore a specific set of configuration settings in the configuration list by defining the index.
Example
SOUR:CONF:LIST:CRE "biasLevel"
SOUR:FUNC VOLT
SENS:FUNC "CURR"
SOUR:VOLT:LEV 5
SOUR:CONF:LIST:STORE "biasLevel"
TRIG:BLOCK:CONF:RECALL 1, "biasLevel", 1
Also see
Configuration lists
[:SENSe[1]]:CONFiguration:LIST:STORe
:SOURce[1]:CONFiguration:LIST:STORe
6-198
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
A string that defines the configuration list to recall
The index in the configuration list to recall; defaults to 1 if not selected
The second source or measure configuration list from which to recall
settings; the type must be opposite of <configurationList>
The index in the second configuration list to recall; defaults to 1 if not
selected
(on page 3-30)
(on page 6-78)
(on page 6-88)
Where saved
Save settings
Create a configuration list named
biasLevel. Set the source function to 5 V
and the measure function to current.
Store the source settings at index 1 in the
configuration list named biasLevel.
Recall index 1 of the configuration list
named biasLevel as block 1 of the trigger
model.
Default value
Not applicable
2461-901-01 A/November 2015