Section 6: SCPI command reference
:SOURce[1]:PULSe:SWEep:<function>:LIST
This command sets up a pulse sweep based on a configuration list, which allows you to customize the sweep.
Type
Command only
Usage
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>"
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>, <failAbort>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>, <failAbort>, <configListName>
<function>
<pulseWidth>
<measEnable>
<bufferName>
<startIndex>
<count>
<delay>
<offTime>
<failAbort>
<configlistName>
6-116
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The source function:
Voltage pulse sweep: VOLTage
Current pulse sweep: CURRent
The time at the amplitude level for each pulse:
Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
Enable or disable measurements at the top of each pulse:
Enable: ON (default)
Disable: OFF
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
The index in the configuration list where the pulse sweep starts; default is 1
The number of pulse sweeps; default is 1:
Infinite loop: 0
Finite loop: 1 to 268435455
The amount of time that the instrument stays at bias level before each pulse (in
seconds): 0 to 10,000 s
The amount of time that the instrument stays at bias level after each pulse (in
seconds): 0 to 10,000 s
Determines if the sweep is stopped immediately if a limit is exceeded; options are:
Abort the sweep if the source bias or pulse limit is exceeded: ON
Complete the sweep if the source limit is exceeded: OFF
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
Where saved
Save settings
Default value
Not applicable