Section 6: SCPI command reference
Details
This command sets the high limit for the limit Y test for the selected measurement function. When limit
Y testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
If you send this command without the <function> parameter, it will set the limit for all measure
functions. It will not change the setting for a digitize function.
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
:CALCulate2:<function>:LIMit<Y>:STATe
6-34
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
(on page 6-30)
(on page 6-32)