smu.digitize.readwithtime()

This function initiates digitize measurements and returns the last actual measurement and time information in
UTC format without using the trigger mode.
Type
TSP-Link accessible
Function
Yes
Usage
reading, seconds, fractional = smu.digitize.readwithtime()
reading, seconds, fractional = smu.digitize.readwithtime(bufferName)
reading
seconds
fractional
bufferName
Details
This command initiates digitize measurements using the present function settings, stores the
readings in a reading buffer, and returns the last reading.
The smu.digitize.count attribute determines how many measurements are performed.
When you use a reading buffer with a command or action that makes multiple readings, all readings
are available in the reading buffer. However, only the last reading is returned as a reading with the
command.
If you define a specific reading buffer, the reading buffer must exist before you make the
measurement.
Example
print(smu.digitize.readwithtime())
Also see
smu.digitize.count
trigger.model.load() — SimpleLoop
2461-901-01 A/November 2015
Affected by
The last reading of the measurement process
Seconds in UTC format
Fractional seconds
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer; if no buffer is specified, this parameter
defaults to defbuffer1
(on page 8-107)
(on page 8-270)
Where saved
Print the last digitize measurement and time information in
UTC format, which will look similar to:
-0.0003882925875
Section 8: TSP command reference
Default value
1415795836
0.946164546
8-125