Section 9: Frequently asked questions (FAQs)
Model 2461 Interactive SourceMeter® Instrument Reference Manual
You cannot combine the command sets.
As delivered from Keithley Instruments, the Model 2461 is set to work with the Model 2461 SCPI
command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select the button next to Command Set.
4. Select the command set.
5. You are prompted to reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
9-6
2461-901-01 A/November 2015