OPERATION
3.6.1 Trigger Source
The trigger source is selected by pressing the SOURCE key.
The instrument
will display one of the trigger sources, as
summarized in Table 3-3. To select a different source, press
uprange or downrange
repeatedly until the desired trig-
ger source is displayed.
The new trigger source is saved
once the source is displayed.
To return to the previous
mode, press SOURCE.
Note that the SOURCE key is inoperative when the instru-
ment is in the dual-channel
display mode.
NOTE
When you select external, input signal, or other
channel
triggering,
you must first arm the AID
converter by pressing the TRIGGER button.
The
measurement
sequence will then begin when the
appropriate
trigger stimulus
occurs.
Table 3-3. Trigger Source Display Messages
Message
Selected Trigger Source
LMMEDIATE
TRlGGER
button or IEEE trig-
gcr (X, GET, etc.)
EXTERNAL
Pulse applied to TRIGGER
IN
jack.
INPUT SIGNAL
Input signal, depending
on
selected slope and levels.
OTHER CHANNEL
Second channel when 1944 op-
tion installed.
TRIG ON TALK
IEEE talk command
TRIG ON X
IEEE X command
TRIG ON GET
IEEE GET command
3.62 Single/Continuous
Arming
Modes
A/D arming can be set up to operate in two different ways:
in continuous
or single modes. In a continuous
mode, the
unit will automatically
rearm the A/D converter after each
measurement
sequence.
The unit
will
then
perform
another measurement
sequence when it receives the ap-
propriate trigger stimulus (as selected with the SOURCE
key). In the single mode, however, the A/D converter must
be manually
re-armed after each measurement
(the A/D
converter can be re-armed from the front panel by press-
ing TRIGGER).
Single/continuous
arming selection is performed by press-
ing the SGLiCONT
button to select the desired mode; the
appropriate
indicator will then display the selected mode.
3-10
When the instrument
is in the dual-channel
display mode,
the SGLiCONT
button
and associated indicators
are
inoperative.
NOTES:
I. If the instrument
is in the immediate (TRIGGER button),
continuous
mode, it will process measurements
con-
tinuously
without
requiring
additional
triggers.
2. If the selected trigger source is the input signal, exter-
nal trigger, or other channel,
the unit will require
a
separate
trigger
stimulus
for
each measurement
regardless of whether
the unit is in the single or con-
tinuous
armed modes.
3. When going from the single, immediate
mode to the
continuous,
immediate
mode, it will be necessary fol
you to press the TRIGGER button to begin processing
measurements.
4. The decimal point in the display (between second and
third digits from the right) will turn on to indicate that
the A/D converter is armed and waiting for a trigger.
Discussion
Basically,
there are two steps needed for initiating
a
measurement
sequence. First, the A/D converter must be
armed, so that it is running
and processing data. Second-
ly, the unit must be triggered by the appropriate
trigger
stimulus (determined
by the selected source) before it will
perform a measurement
sequence. The only exception to
this two-step process is when the immediate, front panel
mode is selected. In this situation, pressing TRIGGER per-
forms both steps simultaneously
so that the A/D converter
is armed and the measurement
is initiated
with a single
action.
There are two fundamental
reasons why the A/D converter
must be running
before a measurement
sequence takes
place. First, when input signal triggering
is in effect, the
A/D converter must supply input signal amplitude
data
to the digital comparator circuits so that they can trigger
a measurement
when the programmed
trigger threshold
is reached. Secondly, when using pre- or post-trigger, the
unit must continuously
store data so that it can later flag
the beginning
of the measurement
relative to the pro-
grammed trigger point.
3.6.3 TRIGGER
Button
Operation
The Model 194 can be triggered by the front panel TRIG-
GER button, if the immediate mode has been selected as
the trigger source (paragraph 3.6.1). Once activated, press-
ing TRIGGER will initiate a single or continuous
sequence
of measurements,
depending
which of those two modes
was selected with the SGLiCONT
button.