Smu.measure.userdelay[N] - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 8: TSP command reference

smu.measure.userdelay[N]

This attribute sets a user-defined delay that you can use in the trigger model.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
delayTime = smu.measure.userdelay[N]
smu.measure.userdelay[N] = delayTime
delayTime
N
Details
To use this command in a trigger model, assign the delay to the dynamic delay block.
The delay is specific to the selected function.
Example
smu.measure.userdelay[1] = 5
trigger.model.setblock(1, trigger.BLOCK_SOURCE_OUTPUT, smu.ON)
trigger.model.setblock(2, trigger.BLOCK_DELAY_DYNAMIC,
trigger.USER_DELAY_M1)
trigger.model.setblock(3, trigger.BLOCK_MEASURE)
trigger.model.setblock(4, trigger.BLOCK_SOURCE_OUTPUT, smu.OFF)
trigger.model.setblock(5, trigger.BLOCK_BRANCH_COUNTER, 10, 1)
trigger.model.initiate()
Set user delay for measure 1 to 5 s.
Set trigger block 1 to turn the source output on.
Set trigger block 2 to a dynamic delay that calls measure user delay 1.
Set trigger block 3 to make a measurement.
Set trigger block 4 to turn the source output off.
Set trigger block 5 to branch to block 1 ten times.
Start the trigger model.
Also see
trigger.model.setblock() — trigger.BLOCK_DELAY_DYNAMIC
8-174
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
The delay (0 for no delay, or 167 ns to 10 ks)
The user delay to which this time applies (1 to 5)
Where saved
Configuration script
Measure configuration list
(on page 8-288)
Default value
0