Model 2461 Interactive SourceMeter® Instrument Reference Manual
5. Set the parameters for the sweep command.
6. Set the measurement function using [:SENSe[1]]:FUNCtion.
7. Set the measurement range using [:SENSe[1]]:<function>:RANGe[:UPPer].
8. Make any other settings appropriate to your sweep.
9. Send :INITiate to start the sweep.
To save your settings, save them to a user-saved setup using the *SAV command.
For example sweeps, see
For detail on the commands and options listed above, see the following command descriptions:
[:SENSe[1]]:FUNCtion[:ON]
[:SENSe[1]]:<function>:RANGe[:UPPer]
:SOURce[1]:FUNCtion[:MODE]
:SOURce[1]:<function>:RANGe
:SOURce[1]:CONFiguration:LIST:CREate
:SOURce[1]:CONFiguration:LIST:STORe
:SOURce[1]:SWEep:<function>:LINear
:SOURce[1]:SWEep:<function>:LINear:STEP
:SOURce[1]:SWEep:<function>:LIST
:SOURce[1]:SWEep:<function>:LOG
Setting up a sweep using TSP commands
To set up a sweep using TSP commands, you send one of the following commands:
smu.source.sweeplinear(): Sets up a linear sweep for a fixed number of measurement
points.
smu.source.sweeplinearstep(): Sets up a linear source sweep configuration list and
trigger model with a fixed number of steps.
smu.source.sweeplist(): Sets up a sweep based on a configuration list, which allows you to
customize the sweep.
smu.source.sweeplog(): Sets up a logarithmic sweep for a set number of measurement
points.
2461-901-01 A/November 2015
Sweep programming examples
(on page 6-81)
(on page 6-65)
(on page 6-95)
(on page 6-97)
(on page 6-83)
(on page 6-88)
(on page 6-124)
(on page 6-126)
(on page 6-128)
(on page 6-130)
Section 3: Functions and features
(on page 3-61).
3-59