Acal:lastrun:time; Acal:run - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference

:ACAL:LASTrun:TIME?

This command returns the date and time when autocalibration was last run.
Type
Query only
Usage
:ACAL:LASTrun:TIME?
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
MM/DD/YYYY is the month, date, and year
HH:MM:SS.NNNNNNNNN is the hour, minute, second, and fractional second
Example
ACAL:LAST:TIME?
Also see

:ACAL:RUN

(on page 6-18)
:ACAL:RUN
This command immediately runs autocalibration and stores the constants.
Type
Command only
Usage
:ACAL:RUN
Details
During autocalibration, all necessary autocalibration steps are completed, calibration constants are
updated, and autocalibration last run information is updated.
When an autocalibration command is received, the instrument runs the autocalibration to completion
before executing the next command.
Example
ACAL:RUN
Also see
Autocalibration
6-18
Affected by
Not applicable
Affected by
Not applicable
(on page 3-51)
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Where saved
Not applicable
Returns the date and time when auto calibration
was last run.
Example output:
08/11/2014 16:30:26.745369595
Where saved
Nonvolatile memory
Auto calibration starts running.
Default value
Not applicable
Default value
Not applicable
2461-901-01 A/November 2015