Using Model 2461 Tsp-Link Commands With Other Tsp-Link Products - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
TSP-Link synchronization programming example
The programming example below illustrates how to set bit B1 of the TSP-Link digital I/O port high,
and then read the entire port value:
tsplink.line[1].mode = tsplink.MODE_DIGITAL_OPEN_DRAIN
-- Set bit B1 high.
tsplink.line[1].state = 1
-- Read I/O port.
data = tsplink.readport()
print(data)
The output would be similar to:
7
To read bit B1 only:
-- To read bit B1 only
data = tsplink.line[1].state
print(data)
The output would be similar to:
tsplink.STATE_HIGH

Using Model 2461 TSP-Link commands with other TSP-Link products

If you are connecting the Model 2461 in a system with other TSP-Link products, be aware that some
of the TSP-Link commands may be different. You can use the earlier versions of the commands, but
be aware that they may not be supported in future versions of the product.
Commands that are the same in all TSP-Link products:
tsplink.group
tsplink.master
tsplink.node
tsplink.readport()
tsplink.state
tsplink.writeport()
2461-901-01 A/November 2015