System:eventlog:save; System:gpib:address - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference

:SYSTem:EVENtlog:SAVE

This command saves the event log to a file on a USB flash drive.
Type
Command only
Usage
:SYSTem:EVENtlog:SAVE "<filename>"
:SYSTem:EVENtlog:SAVE "<filename>", <eventType>
<filename>
<eventType>
Details
This command saves all event log entries to a USB flash drive.
If you do not define an event type, the instrument saves all event log entries.
The extension .csv is automatically added to the file name.
Example
SYST:EVEN:SAVE "/usb1/July_error_log", ERR
Also see
:SYSTem:CLEar

:SYSTem:GPIB:ADDRess

This command contains the GPIB address.
Type
Command and query
Usage
:SYSTem:GPIB:ADDRess <n>
:SYSTem:GPIB:ADDRess?
<n>
6-152
Affected by
Not applicable
A string that holds the name of the file to be saved
Limits the event log entries that are saved to specific types; set to:
ERRor: Saves only error entries
WARNing: Saves only warning entries
INFormational: Saves only informational entries
ALL: Saves all event log entries (default)
(on page 6-145)
Affected by
Not applicable
The GPIB address of the instrument (1 to 30)
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Where saved
Not applicable
Saves the error events in the event log to a
file on the USB flash drive named
July_error_log.csv.
Where saved
Nonvolatile memory
Default value
Not applicable
Default value
18
2461-901-01 A/November 2015