Keithley 2461 Reference Manual page 233

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 3: Functions and features
Using TSP commands
reset()
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.autorange = smu.ON
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.source.ilimit.level = 0.02
smu.source.sweeplinear("RES", 0, 10, 21, 200e-3)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 21, defbuffer1.sourcevalues, defbuffer1.readings)
Logarithmic sweep with a current source
The following examples perform a logarithmic sweep using a current source. They perform the
following actions:
Reset the instrument to its defaults.
Set the source function to current.
Set the source range to 100 mA.
Set up a logarithmic sweep from 100 µA to 100 mA in 10 steps with a source delay of 10 ms, a
sweep count of 1, and a fixed source range. In TSP only, name the configuration list that is
created for this sweep RES.
Set the measure function to current.
Set the current range to 100 µA.
Start the sweep.
No buffer is defined, so the data is stored in defbuffer1. See
more information on reading buffers.
Using SCPI commands
*RST
SOUR:FUNC CURR
SOUR:CURR:RANG 100e-3
SOUR:CURR:VLIM 20
SENS:FUNC "VOLT"
SENS:VOLT:RANG 20
SOUR:SWE:CURR:LOG 100e-6, 100e-3, 10, 10e-3
INIT
*WAI
TRAC:DATA? 1, 10, "defbuffer1", SOUR, READ
Using TSP commands
reset()
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.range = 100e-3
smu.source.vlimit.level = 20
smu.source.sweeplog("RES", 100e-6, 100e-3, 10, 10e-3)
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.range = 20
trigger.model.initiate()
waitcomplete()
printbuffer(1, 10, defbuffer1.sourcevalues, defbuffer1.readings)
3-62
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Reading buffers
(on page 3-2) for
2461-901-01 A/November 2015