Section 6: SCPI command reference
:CALCulate2:<function>:LIMit<Y>:STATe
This command enables or disables a limit test on the measurement from the selected measure function.
Type
Command and query
Usage
:CALCulate2:<function>:LIMit<Y>:STATe <state>
:CALCulate2:<function>:LIMit<Y>:STATe?
<function>
<Y>
<state>
Details
This command enables or disables a limit test for the selected measurement function. When this
attribute is enabled, the limit Y testing occurs on each measurement made by the instrument. Limit Y
testing compares the measurements to the high and low limit values. If a measurement falls outside
these limits, the test fails.
If you send this command without the <function> parameter, it sets the state of math operations
for all math functions. It will not change the setting for a digitize function.
6-32
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measure function:
Current: CURRent[:DC]
•
Resistance: RESistance
•
Voltage: VOLTage[:DC]
•
The digitize function:
Digitize current: DIGitize:CURRent
•
Digitize voltage: DIGitize:VOLTage
•
Limit number: 1 or 2
Disable the limit test: OFF or 0
Enable the limit test: ON or 1
Where saved
Save settings
Measure configuration list
Default value
0 (OFF)
2461-901-01 A/November 2015