:TRIGger:BLOCk:BUFFer:CLEar

This command defines a trigger model block that clears the reading buffer.
Type
Command only
Usage
:TRIGger:BLOCk:BUFFer:CLEar <blockNumber>
:TRIGger:BLOCk:BUFFer:CLEar <blockNumber>, "<bufferName>"
<blockNumber>
<bufferName>
Details
When trigger model execution reaches the buffer clear trigger block, the instrument empties the
specified reading buffer. The specified buffer can be the default buffer or a buffer that you defined.
If you are clearing a user-defined reading buffer, you must create the buffer before you define this
block.
Example
TRIG:LOAD "EMPTY"
TRIG:BLOC:BUFF:CLE 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
Also see
Buffer clear block
:TRACe:MAKE
2461-901-01 A/November 2015
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The name of the buffer, which must be an existing buffer; if no buffer is
defined, defbuffer1 is used
(on page 3-108)
(on page 6-165)
Section 6: SCPI command reference
Where saved
Save settings
Reset trigger model settings.
Clear defbuffer1 at the beginning of the
trigger model.
Loop and take 5 readings.
Delay 1 s.
Loop three more times back to block 2.
At end of execution, 15 readings are stored in
defbuffer1.
Default value
Not applicable