To Get Measurement Trace Using Ascii Format; Set The Receive Array; Sample Program : To Get Measurement Trace Using Ascii Format - Agilent Technologies 4396B Gpib Programming Manual

Network/spectrum/impedance analyzer
Hide thumbs Also See for 4396B:
Table of Contents

Advertisement

To Get Measurement Trace Using ASCII Format

d
c
Figure 4-2. Sample Program : To Get Measurement Trace Using ASCII Format

Set the Receive Array

Line 180 sets the array size to the analyzer's maximum number of measurement points (801).
In this example, it is assumed that the analyzer is in the network analyzer mode of operation,
in which each point has complex data. If you use the analyzer in the spectrum analyzer mode,
each measurement point has only real data, so you must set the data array
If the number of measurement points changes, then so does the number of data. You must
control the number of entered measurement data (see lines 210 and 230).
4-4
Reading Measurement Data
a
b
as follows: