Making Measurements
Using Markers

To Use Polar Format Markers

The analyzer can display the marker value as magnitude and phase, or as a
real/imaginary pair:
magnitude and phase,
You can use these markers only when you are viewing a polar display format. (The format
is available from the
For greater accuracy when using markers in the polar format, it is
NOTE
recommended to activate the discrete marker mode. Press
MKR MODE MENU
1. To access the polar markers, press:
POLAR
Format
2. Select the type of polar marker you want from the following choices:
• Choose
LIN MKR
marker. The magnitude values appear in units and the phase values appear in
degrees.
• Choose
LOG MKR
the active marker. The magnitude values appear in dB and the phase values appear
in degrees.
• Choose
Re/Im MKR
complex data is separated into its real part and imaginary part. The analyzer shows
the real part as the first marker value (M cos Θ), and the second value is the
imaginary part (M sin Θ, where M = magnitude).
Figure 1-20 Example of a Log Marker in Polar Format
1-32
gives linear magnitude and phase,
LIN MKR
gives the real value first, then the imaginary value.
Re/Im
key.)
Format
MARKERS:DISCRETE
MARKER MODE MENU
Marker Fctn
if you want to view the magnitude and the phase of the active
if you want to view the logarithmic magnitude and the phase of
if you want to view the real and imaginary pair, where the
LOG MKR
Marker Fctn
.
POLAR MKR MENU
gives log