Table B-2. Averaging Times; Measurements - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands

Measurements

Measurements
B-8
Model 48i Trace Level-Enhanced Instruction Manual
Command
Description
srec per
Reports/sets short record logging period
stream per
Reports/sets current set time interval for streaming data
stream time
Reports/sets a time stamp to streaming data or not
temp comp
Reports/sets temperature compensation on or off
time
Reports/sets current time (24-hour time)
up
Simulates pressing up pushbutton
zero
Sets zero/span valves to zero mode
avg time
high avg time
low avg time
These commands report the averaging time in seconds when operating in
single range, or averaging time used with the high and low ranges when
operating in dual or auto range mode. The example below shows that the
averaging time is 300 seconds, according to
Send:
avg time
Receive:
avg time 11:300 sec
set avg time selection
set high avg time selection
set low avg time selection
These commands set the averaging time, high and low averaging times,
according to
Table
B–2. The example below sets the low range averaging
time to 120 seconds.
Send:
set low avg time 8
Receive:
set low avg time 8 ok
Table B–2. Averaging Times
Selection
Averaging Time (seconds)
0
1 seconds
1
2
2
5
3
10
4
20
5
30
6
60
7
90
Table