Table B-3. Averaging Times - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Measurements
Measurements
B-10
Model 48i Instruction Manual
Command
Description
version
Reports version of all the firmware components
zero
Sets zero/span valves to zero mode
zero cal reset
Reports/sets zero cal reset on/off
zero calibrate
Sets CO coefficient
zero dev
Reports/sets zero deviation (maximum zero check offset)
zero dur
Reports/sets how long zero gas is sampled by the instrument
zs avg time
Reports/sets zero/span averaging time
zs period
Reports/sets zero/span period
avg time
high avg time
low avg time
These commands report the averaging time in seconds when operating in
single range, or averaging time used with the high and low ranges when
operating in dual or auto range mode. The following example shows that
the averaging time is 300 seconds, according to
Send:
avg time
Receive:
avg time 11:300 sec
set avg time selection
set high avg time selection
set low avg time selection
These commands set the averaging time, high and low averaging times,
according to
Table
B–3. The following example sets the low range
averaging time to 120 seconds.
Send:
set low avg time 8
Receive:
set low avg time 8 ok
Table B–3. Averaging Times
Averaging Time (seconds)
Selection
0
1 seconds
1
2
2
5
3
10
4
20
5
30
6
60
Table
B–3.
Thermo Fisher Scientific
Page
B-19
B-37
B-31
B-26
B-32
B-32
B-33
B-32