3.2.4 Temperature Calibration or
Manual Temperature Input
3.2.4a Temperature Calibration
When a conductivity electrode with built-in ATC sensor is
connected, use the Temperature Calibration menu to
perform a one-point temperature offset calibration, up to
±5.0 °C.
Press the or key to scroll through the Setup
Menu list until the Temp. Calibration item is shown.
The temperature value will blink, indicating the
reading from the ATC probe is being measured.
Once the reading is stable, the value will stop
blinking and the checkmark icon will be shown.
Press the or key to adjust the temperature
value in the 3
rd
field. The temperature offset value
will be shown in the 2
a. Press the key to increase the value by 0.1°.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.1°.
Press and hold the key to quickly decrease.
Press the "Enter" key to save the change.
16
Chapter 3 | Setup Menu
nd
field.
3.2.4b Manual Temperature Input
When a conductivity electrode without ATC sensor is
connected, use the Manual Temperature Input menu to
enter the sample temperature value, from -5.0 °C to
105.0 °C.
Press the or key to scroll through the Setup
Menu list until the Manual Temperature Input item is
shown.
Press the or key to adjust the temperature
value.
a. Press the key to increase the value by 0.1°.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.1°.
Press and hold the key to quickly decrease.
Press the "Enter" key to save the change.