Analog Input Calibration; Analog Input Calibrate Zero - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Analog Input
Calibration
Analog Input Calibrate
Thermo Fisher Scientific
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
The Analog Input Calibration menu is used to calibrate the 8 analog input
channels at both zero and full-scale. The Analog Input Calibration menu is
visible only when the I/O expansion board is installed. The analog input
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see
the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Cal.
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
RANGE
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
Zero
In the Main Menu, choose Service > Analog Input Cal > select Channel
> Calibrate Zero.
SPAN
3697

INC/DEC
10 V
AVG
DIAGS ALARM
AVG
DIAGS ALARM
AVG
DIAGS ALARM
V1
"Service
Mode" earlier in
Model 48i Instruction Manual
Operation
Service Menu
3-87