Analog Input Calibration

Analog Input Calibrate
Thermo Fisher Scientific
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
The Analog Input Calibration menu is a selection of 8 analog inputs (if the
I/O expansion board option is installed) to calibrate, and allows the user to
select the calibration action zero or span. The analog input calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Calibration.
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
RANGE
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
Zero
SPAN
3397
INC/DEC
10 V
AVG
DIAGS ALARM
"Service
Mode" earlier in the chapter.
AVG
DIAGS ALARM
AVG
DIAGS ALARM
Model 42i Trace Level Instruction Manual
Service Menu
V1
Operation
3-83