Analyzer Hardware Configuration and Calibration
Section 5
be allowed to complete and is timed at 60 seconds to allow the user to enter the volume
collected in 60 seconds (or 1 minute). If the measured volume amount is not entered within 90
seconds the pump calibration procedure is aborted.
Upon successful completion of the pump calibration procedure, the user has the choice to
calibrate the remaining pumps or to edit the calibrated values.

Analog Outputs Calibration

The analyzer has two isolated analog outputs. The outputs are rated 0/4-20mA with a maximum
load of 900 ohms. They can be independently calibrated.
From the MAIN MENU, follow the sequence:
MENU ➔ SERVICES ➔ HARDWARE SETUP ➔ CALIBRATE ANALOG
OUTPUTS
The user is asked to connect a 900 ohms resistor to each two analog outputs. The calibration
requires the use of a current meter (0–100 mA range) placed in series with the output under
test.
The user may select one output at a time for calibration. The calibration starts by adjusting the
full scale to 20mA. The -/+ buttons are used to adjust the output to 20mA. The instrument waits
for the user to confirm the calibration by pressing ENTER. If there is no user response within 90
seconds, the analog output procedure calibration is aborted.
Upon successful calibration of one of the analog outputs, the user has the choice of calibrating
the second analog output.
The calibrated values for the analog outputs are not available for editing. For testing the
calibrated outputs use:
MENU ➔ SERVICES ➔ DIAGNOSTICS ➔ ANALOG OUTPUT ➔ CHANNEL
–1:
By pressing ENTER the user may select the signal level from the following settings:
0, 4, 10, 12, and 20 mA
2230XP Silica Analyzer
Thermo Scientific Orion
37