Analog Output Full Scale
Thermo Scientific
Use the following procedure to calibrate the output channel to full scale.
Calibration
This analog output calibration procedure reflects the full scale calibration
for analog output voltage for demonstration purposes. To calibrate the 20
mA current calibration, follow the same procedure, by selecting the 20 mA
current calibration option.
Note This adjustment should only be performed by an instrument service
technician.
1. From the Home screen, choose Settings>Communications>Analog
I/O>Analog Output Calibration.
2. Depending on the output type being used, select either Analog Out
Full Scale Calibration (Voltage) or Analog Out 20.000 mA Calibration
(Current).
3. A confirmation screen is presented. Select Continue to proceed with
the calibration or Return to Previous Screen.
L
Optional Equipment
Connecting External Devices
9-9
43iQTL Instruction Manual