Analog Output Calibrate Zero; Analog Output Calibrate Full-Scale - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Service Menu
Analog Output Calibrate
Analog Output Calibrate
3-86
Model 48i Instruction Manual
In the Main Menu, choose Service > Analog Out Cal.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a volt
Zero
meter to the output and adjust the output until it reads 0.0 V on the meter
for a voltage channel, or either 0 or 4 mA for a current channel (depending
on the selected range). See the "set output to:" field on the display.
In the Main Menu, choose Service > Analog Out Cal > select Channel
> Calibrate Zero.
ANALOG OUTPUT CAL:
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
Full-Scale
a volt meter to the output and adjust the output until the meter reads the
value shown in the "set output to:" field, in either V or mA, depending on
the selected output channel.
In the Main Menu, choose Service > Analog Out Cal > select Channel
> Calibrate Full Scale.
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE
AVG
DIAGS ALARM
CALIBRATE FULL SCALE
RANGE
AVG
DIAGS ALARM
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
DIAGS ALARM
ZERO
V1
100

INC/DEC
0.0 V