Analog Output Calibrate Zero; Analog Output Calibrate Full-Scale - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Operation
Service Menu
Analog Output Calibrate
Analog Output Calibrate
3-82
Model 42i Trace Level Instruction Manual
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a meter
Zero
to the output and adjust the output until it reads 0.0 V on the meter.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Zero.
ANALOG OUTPUT CAL:
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
Full-Scale
a meter to the output and adjust output until it reads the value shown in
the set output to: field.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Full Scale.
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
RANGE
AVG
DIAGS ALARM
CALIBRATE FULL SCALE
RANGE
AVG
DIAGS ALARM
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
DIAGS ALARM
ZERO
V1
100
INC/DEC
0.0 V
Thermo Fisher Scientific