Calibrate Pressure Zero; Calibrate Pressure Span; Restore Default Pressure Calibration - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Service Menu

Calibrate Pressure Span

Restore Default Pressure
3-80
Model 48i Instruction Manual
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
Note Wait at least 30 seconds for the reading to stabilize before saving the
value.
In the Main Menu, choose Service > Pressure Calibration > Span.
The Restore Default Pressure Calibration screen allows the user to reset the
pressure calibration configuration values to factory defaults.
Calibration
In the Main Menu, choose Service > Pressure Calibration > Set
Defaults.
RESTORE DEFAULT CAL:
RANGE
AVG
DIAGS ALARM

CALIBRATE PRESSURE ZERO:

CURRENTLY:
SET TO:
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE
AVG
DIAGS ALARM
CALIBRATE PRESSURE SPAN:
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
RESTORE
753.0 mmHg
0.0 mmHg ?
753.0 mmHg
760.0
mmHg ?
SAVE
RESTORE DEFAULT CAL:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
RESTORE
TO?x
DIAGS ALARM