The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
In the Main Menu, choose Service > Pressure Calibration > Span.
The Restore Default Calibration screen allows the user to reset the pressure
calibration configuration values to factory defaults.
Calibration
In the Main Menu, choose Service > Pressure Calibration > Set
Defaults.

CALIBRATE PRESSURE ZERO:

CURRENTLY:
SET TO:
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE
AVG
DIAGS ALARM
CALIBRATE PRESSURE SPAN:
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
232.5 mmHg
0.0 mmHg
232.0 mmHg
760.0
mmHg ?
SAVE