Pressure Calibration; Calibrate Pressure Zero - Thermo Scientific 48i Instruction Manual

Trace level-enhanced gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Service Menu

Pressure Calibration

Calibrate Pressure Zero

3-72
Model 48i Trace Level-Enhanced Instruction Manual
The Pressure Calibration submenu is used to calibrate the pressure sensor
to zero, span, or restore factory default values. The pressure calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see
The pressure sensor's zero counts and span slope are displayed on the
menu.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Pressure Calibration.
PRESSURE SENSOR CAL:
>ZERO
SPAN
SET DEFAULTS
RANGE
AVG
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
In the Main Menu, choose Service > Pressure Calibration > Zero.
CALIBRATE PRESSURE ZERO:
CURRENTLY:
SET TO:
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE
AVG
"Service
Mode" earlier in this chapter.
72
1.1416
DIAGS ALARM
10.0 mmHg
0.0 mmHg
DIAGS ALARM