Terms Definition - Thermo Scientific 2210AM Manual

Ammonia analyzer
Table of Contents

Advertisement

Terms Definition

Terms
ALARM
Type
Value
Hysteresis
RELAY
MODE
ANALOG OUTPUT
Type
Low
High
EXTERNAL INPUT
Label
Event
Action
DISPLAY PARAMETERS
Contrast
Back Light On-Time
Display Help
Display Info
Results Decimal Point
Large Fonts Screen
Orion 2210AM Ammonia Analyzer
Description
Assigning a fault type to the alarm.
Boundary value for triggering an alarm,
only for low limit and high limit type.
Value gap or difference before an event
change.
Up to FOUR spare relays available.
Allows fault ON or OFF operation.
Normal or reverse acting.
Low concentration value.
High concentration value.
User programmable label.
Dry contact event.
Selected function.
To adjust the display contrast.
Screen saver time duration for backlight
of display.
Help message for the user to navigate
between the menus.
Message showing the status
of the current cycle performing.
Selectable precision to show the analysis
result.
To display the result with large fonts.
Possible Values, Options
Low Limit, High Limit, Fluidics Fault,
General Fault, not Assigned
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
parameters
result Decimal Point)
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
parameters
result Decimal Point
Normal or Fail Safe
0 to 20, 4–20, 20-0 or 20–4
Close/Open or Open/Close
Analyze Once, Run/Stop, Calibrate, Validate,
Not assigned
120 to 200
0 to 255 minutes, 0 for Continuously On
Enabled, Disabled
Enabled, Disabled
XXXX.X, XXX.XX, XX.XXX, X.XXX, XXXXX,
Auto
Enabled, Disabled
Terms Definition
Display
Display
|
33