Prime Fluidics Menu And New Reagents Blank; Diagnostics Menu - Thermo Scientific Orion 2230XP User Manual

Silica analyzer
Table of Contents

Advertisement

2230XP Silica Analyzer

Prime Fluidics Menu and New Reagents Blank

Auto Prime All
This menu runs an automatic priming sequence for all reagents, standards and sample based
on the priming volume specified by the user.
Prime Input Source (Sample/Reagent/...)
Allows technician to selectively prime from the available list of devices.
Priming Volume
Allows technician to change the volume used for priming.
Press + to Navigate to NEXT Page
The new reagents blank is the ZERO intercept or offset. A user entry here will affect the zero
calibration value of the calibration parameters.
Set NEW reagents BLANK
Allows the operator to set the BLANK value for the NEW reagents.This value is usually supplied
with the new batch of reagents and its value may vary between 0 to -10. Default value is -1.0.
PLEASE NOTE THE (-) SIGN.

Diagnostics Menu

Pumps, Valves and Stirrer
Allows technician to selectively check the functionality of the pumps, valves and stirrer.
Inputs Status
Allows technician to check the status of the external inputs. Once a function has been assigned
to a digital input, it disappears from the next menu of available functions for the next input.
External Control
The analyzer can be directed to initiate an analysis or calibration cycle or change operating
mode under the command of an external controller using the digital input lines. A command on
the digital input is represented by a contact closure (as the programmed EVENT) that occurs for
more than 1 second. A new command received supersedes the previous command; in essence
it will interrupt the previous command to execute the new command.
Analyze once
RUN/STOP
Calibrate
Validate (the source is defined in the validation menu)
Not assigned
Menu Overview
Section 7
Thermo Scientific Orion
55

Advertisement

Table of Contents
loading

Table of Contents