The Streaming Data protocol transmits user-selected measurement data via
the serial port in real-time for capture by a serial printer, data logger, or PC.
A null modem (crossed) cable is required when connecting the analyzer to
an IBM Compatible PC. However, a straight cable (one to one) may be
required when connecting the analyzer to other remote devices. As a
general rule, when the connector of the host remote device is female, a
straight cable is required and when the connector is male, a null modem
cable is required.
Data Format:
1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200 BAUD
8 data bits
1 stop bit
no parity
All responses are terminated with a carriage return (hex 0D)
Refer to
Table 8–1
Table 8–1. RS-232 DB Connector Pin Configurations
DB9 Pin
Function
2
RX
3
TX
7
RTS
8
CTS
5
Ground
The instrument uses a four wire RS-485 configuration with automatic flow
control (SD). Refer to
Table 8–2. RS-485 DB Connector Pin Configuration
DB9 Pin
Function
2
+ receive
8
- receive
7
+ transmit
3
- transmit
5
ground
for the DB9 connector pin configuration.
Table 8–2
for the DB9 connector pin configuration.