Download Print this page

Lcd Module Replacement; Figure 7-13. Replacing The Lcd Module - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Servicing

LCD Module Replacement

LCD Module
Replacement
7-12
48iQ Instruction Manual
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1. Turn instrument OFF and unplug the power cord.
2. Gripping from the top corners of the front panel and pull outwards.
3. Using 1/4 wrench, unfasten four #4-40 nuts
Figure 7–13. Replacing the LCD Module
4. Remove cover.
5. Unplug LCD cables from backside of board.
6. Pull board off the standoffs.
(Figure
7–13).