Lcd Module Replacement; Figure 7-13. Replacing The Lcd Module - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Servicing

LCD Module Replacement

LCD Module
Replacement
7-12
42iQHL Instruction Manual
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1. Turn instrument OFF and unplug the power cord.
2. Gripping from the top corners of the front panel and pull outwards.
3. Unfasten four nuts.
Figure 7–13. Replacing the LCD Module
4. Remove cover.
5. Unplug LCD cables from backside of board.
6. Pull board off the standoffs.