Analog Output Calibration - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Analog Output Calibration

Thermo Fisher Scientific
Note This adjustment should only be performed by an instrument service
technician.
Note Wait at least 30 seconds for the reading to stabilize before saving the
value.
In the Main Menu, choose Service > Temperature Calibration.
CALIBRATE AMBIENT TEMP:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see
the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Out Cal.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE
AVG
32.3 C
o
032.5
C ?

MOVE CURSOR
SAVE
DIAGS ALARM
DIAGS ALARM
Operation
Service Menu
"Service
Mode" earlier in
Model 42i Instruction Manual
3-83